The data loggers can be used for various applications within the life science and food and beverage industries. When used with our Automarker Sensor, the following application(s) is covered:
Record events during a session (set time markers)
Interchangeable Sensor Design
TrackSense Pro Basic Wireless Data Logger
Automarker Sensor
The AutoMarker Sensor can be used to record events during a session. When started, time markers can be added to the session either by manual activation of the switch placed at the cable end or automatically when connected directly to a PLC using the PLC relay function.
Examples of applications the AutoMarker would be used in are:
- Logging of holding time start/stop from autoclave controller
- Logging of door openings in warehouse storage, stability chambers, freezers, incubators, etc.
- Logging of general failure on critical components during process e.g. pump failure
- Logging emergency shelf cooling activation in freeze dryer
The high quality wireless TrackSense Pro Basic data loggers consist of high resistant stainless steel and offers high-end technology that allows for stable and highly accurate measurements within different thermal processes.
The TrackSense Pro data logger series are equipped with an Interchangeable sensor option, unlocking several measuring parameters:
✓ Temperature
✓ Pressure
✓ Conductivity
✓ Relative humidity
✓ Vacuum
The data loggers can easily be activated and read by the Multi reader station and combined with the Sky module for real time data via wireless communication.
Key Features and Benefits
One logger - multiple sensor options
21 CFR Part 11 Compliant
Real time data with the SKY module
User-friendly validation software
Usable with E-Val Pro thermocouple system
Sensor with this logger configuration:
Automarker Sensor
Working Principle:
Contact Free Switch Sensor
Switch Cable Plug Type:
LEMO
Number of Switches Per Sensor:
1
Length of Switch Sensor Cable:
100 cm
IP Class:
40
Logger with this sensor configuration:
Pro Basic 3G
Operating Temperature:
+5 to +50 °C
Operating Pressure:
0.9 – 1.1 Bar ABS
Operating Humidity:
10-95%
House Material:
316 L Stainless Steel
Diameter:
25 mm
Length:
44 mm
Weight With Battery:
48 Grams
Memory Capacity:
120,000 Data Points / 120,000 Samples
Minimum Sample Rate:
1 Second
Maximum Sample Rate:
24 Hours
Maximum Start Delay:
14 Days
Intrinsically Safe:
Ex II1GD Ex ia IIC T3 Ga, -30 °C ≤ Tamb ≤ +105 °C
Time Accuracy:
± 5 Seconds Per 24 Hours
Battery:
Please Note: If equipment is used in ATEX environment,
special conditions for safe use stated in ATEX certificates
section 17 must be considered.
The ValSuite® Software
ValSuite is the industry leading validation and calibration software, which is highly attributed to the countless reporting options, impenetrable data integrity, user-calibration options and data analysis features.
The software is used to run all Ellab hardware and can combine various measuring systems to the same study. In addition to its wide range of features, ValSuite is also FDA 21 CFR Part 11 compliant and designed in accordance with GAMP guidelines.
Features and Benefits:
✓ FDA 21 CFR Part 11 compliant
✓ Safe and reliable
✓ Windows security option
✓ Compatible with Windows 10
✓ Available in several languages
✓ Suitable for all sorts of applications
✓ Multiple reporting options
And much more…
Highlighted Features
Suitable Applications
Ellab Applications
Ellab provides a wide range of diverse products and solutions for the pharmaceutical, medical and food industries all around the world.
See all applicationsSuitable Accesories
Ellab offers a wide range of customizable fittings and essential accessories for processes and studies within different industries. By using one of our unique fittings, loggers can be applied to hard-to-reach areas, cold spots and other critical measuring points.
Ellab Accesories
Ellab offers a wide range of customizable fittings and essential accessories for processes and studies within different industries.
See all Accesories