The ETI Recorder Plus is a high quality digital temperature and pressure-indicating device designed for displaying and documenting the processing temperature and pressure in retorts. It complies with FDA 21 CFR Part 113, and it is ideal for thermal processing of low-acid canned foods. The ETI Recorder Plus has a digital display that is easy to read, offers an temperature accuracy of +/- 0.1°C and an pressure accuracy of 0.15% full scale (20…120 °C), and does not constitute a potential source of mercury contamination.
Compliant with FDA 21 CFR Part 113
FDA compliant alternative to mercury-in-glass thermometers
Intelligent 6 mm Diameter Stainless Steel Sensor With Two Independent Pt-1000
Sensing Elements and A/D Converters
Environmental:
0 – 90% RH, non-condensing
Ingress Protection Code:
IP67
Output:
RS485
External Power:
7 to 30 V DC
Battery Backup:
Internal, Rechargeable, 8 Hour
Display Update Rate:
1 Second
Memory:
60,000 Data Points
Digital Pressure Sensor
Communication interface:
RS485
Measuring Range:
0 to 4 bar ABS
Supply Voltage:
8…..28 VDC
Accuracy:
0.15% FS (20…..120°C)
True Output Rate:
100 Hz
Resolution:
0.002% FS
Long Term Stability Type:
Range < 2 Bar: 2mBar
Range >2 Bar: 0.2% FS
Electrical Connection:
Binder-Plug 723 (5 Pole)
Storage- /Operating Temperature Range:
20….150°C (Electronics Max. 120°C)
Protection:
IP64 Optional: IP67 or IP68 (with Cable)
CE-Conformity:
EN 61000-6-1 to 6-4
Material in Contact with Media:
Stainless Steel 316L (DIN 1.4435)/Vitron
Weight:
280g
Documented and Traceable Accuracy
The ETI Recorder Plus has an accuracy of +/- 0.1°C / 0.18°F and is tested against an accurate calibrated reference device and is delivered with a NIST traceable (or equivalent) calibration certificate. The ETI Recorder Plus temperature sensor has a unique ID which is shown on the calibration certificate ensuring traceability.
The combination of full compliance with FDA regulations and Ellab’s renowned quality and high reliability makes the ETI Recorder Plus an excellent choice for retort monitoring. Recognizing that product safety, quality and efficiency are critical to our clients, Ellab has added additional features to the ETI Recorder Plus.
Calibration
The factory recalibration procedure is made simple as the sensor holds both calibration constants and ID which are independent from the display. When sending the sensor in for factory calibration, a certificate will be issued which includes all of the information needed to comply with FDA regulations and ensures the reliability and accuracy of the ETI. The calibration date is easily visible on the digital display. The ETI Recorder Plus display automatically shows a warning 28 days before factory recalibration is due.
The ETI Recorder Plus offers complete process control and monitoring via optional connection to a PLC using Modbus communication.
The temperature, pressure and time measurements are communicated directly to the PLC. This makes it possible for the user to control the process and the machine (autoclave). Monitoring one or several ETI Recorder Plus units is possible when the PLC is connected to a SCADA system. This ensures an accurate overview of the measurements from all ETI Recorder Plus units installed.
To ensure that your process continues to provide safe and sterile results, your autoclave must comply with established standards and norms (EN 285 and ISO 17665). In order to successfully do this, autoclaves require periodic qualification and validation to prove that they sterilize within their qualified parameters – and thereby provide completely clean and sterile materials and equipment.
Retort Sterilization in the food industry refers to any closed vessel or other equipment used for the thermal processing of foods. Typically, the sterilization temperatures vary from 110 to 135°C.
Both static and rotary steam/air retorts, and water immersion/cascade/spray retorts require temperature validation.