NEW
Temperature Range of
-80 to +250 °C
Ellab Temperature Standard
ETS10
The traceable ETS allows for extremely accurate temperature measurements by converting resistance into temperature using its built-in algorithms.
When connected to the ValSuite software, the LiquiCal bath and ETS provide a perfect automatic setup which – saving time and allowing for convenient calibrations.
The ETS has a built-in digital display to indicate the readings in real-time, and can be powered through either USB or LAN (PoE) depending on user-preference.
Key Features and Benefits
Temperature Range -80 to
+250 °C
Accuracy: ±0.015 °C / ±0.025 °C
Separated electronics from heat zone
Robust and versatile design
USB and LAN communication option
Holder to effortlessly mount on walls or steel surfaces
In-built digital display for live readings
ETS10
Housing Material:
Recyclable ABS Plastic
Housing IP Class:
41
Display:
High resolution 3-digit digital display
Interface & Power Source:
USB or Power over Ethernet (PoE)
Temperature Range:
-80 to +250 °C
Accuracy – Including Instrument, PRT and Drift:
-50 to +150 °C:
±0.015 °C
Other temperatures:
±0.025 °C
Measuring Sample Rate:
1 Sec.
Digital Filter:
0-10 Sec.
Resistance:
100 Ω
Sensor Response Time:
T-63%:
33 Seconds
T-90%:
45 Seconds
Sensor Material:
Inconel™ 600
Dimension of Sensor:
ø 6.35 x 300 mm
Minimum Immersion Depth:
100 mm
Environmental:
5 – 50 °C / 0 – 90% RH, non-condensing
Calibration:
Traceable to Teknologisk Institut, Denmark
Calibration Algoritms:
ITS-90
Recommended Calibration Interval:
12 months
Weight:
305 g