NEW
Temperature Range of
-95 to +150 °C
Ellab Temperature Standard
Angled ETS25 - A213
ETS25-A temperature standards are ideal for working with our line of dry blocks to calibrate wireless data loggers and industrial sensors that take up additional space over the insert/calibration zone. When placed in the center of insert , the ETS25-A213 will simply clear the top of the dry block by bending away, thereby providing more room for the sensors undergoing tests.
The traceable ETS allows for extremely accurate temperature measurements by converting resistance into temperature using its built-in algorithms.
When connected to the ValSuite software, the LiquiCal bath and ETS provide a perfect automatic setup which – saving time and allowing for convenient calibrations.
The ETS has a built-in digital display to indicate the readings in real-time, and can be powered through either USB or LAN (PoE) depending on user-preference.
Key Features and Benefits
Temperature Range -95 to
+150 °C
Accuracy ±0.025 °C
Separated electronics from heat zone
Robust and Versatile Design
USB and LAN communication option
Holder to effortlessly mount on walls or steel surfaces
In-built digital display for live readings
ETS25 – A213
Housing Material:
Recyclable ABS Plastic
Housing IP Class:
41
Display:
High resolution 3-digit digital display
Interface & Power Source:
USB or Power over Ethernet (PoE)
Temperature Range:
-95 to +150 °C
Accuracy – Includes Instrument, PRT and Drift:
-95 to +150 °C:
±0.025 °C
Measuring Sample Rate:
1 Sec.
Digital Filter:
0-10 Sec.
Resistance:
100 Ω
Sensor Response Time:
T-63%:
33 Seconds
T-90%:
45 Seconds
Sensor Material:
Inconel™ 600
Sensor Diameter:
ø 6.35 mm
Immersion Depth (vertical length before bend):
190.5 mm
Total Length:
213.4 mm
Environmental:
5 – 50 °C / 0 – 90% RH, non-condensing
Calibration:
Traceable to Teknologisk Institut, Denmark
Calibration Algoritms:
ITS-90
Recommended Calibration Interval:
12 months
Weight:
305 g