The data loggers can be used for various applications within the life science and food & beverage industries.
Differential Pressure Range of +/-62Pa
Capable of CO2 and Air Flow measurement
EMS Compatible
Live Data
Hanwell IceSpy IN-DP-62 Transmitter-Sensor
Differential Pressure Transmitter-Sensor
The Hanwell IceSpy IN-DP-62 wireless transmitter-sensor offers a dual channel multi-purpose analogue transmitter-sensor for monitoring differential pressure.
Instrument Measuring Range +/-62Pa
Compatible with our wireless EMS monitoring system for real-time 24/7 data (anytime and anywhere) and instant alarm notifications.
Key Features and Benefits
Superior performance hardware & high accuracy sensors
Waterproof (relates to Hanwell IceSpy transmitter, sensors used with the transmitter may vary).
All transmitters incorporate a wall bracket
Easily accessible battery for replacement by the user, when required
Low power radio for long distance transmission
Up to 1.5-year battery life
Complies with UKCA, RoHS, EU directives and WEEE
Carries CE Marking
Instrumentation Specification
Dimension (Excl. ancillaries):
130 x 65 x 35mm (40mm including mounting bracket)
5.12 x 2.56 x 1.38 inches (1.57 inches including mounting bracket)
Weight:
150g (5.29 Ounces) (including battery)
Power supply:
1 x 1.5V AA Lithium battery
Case Material:
ABS
Memory capacity:
On board logging capability for 5 days
Input type:
2 Channels 4-20mA
Resolution:
12 BIT
Transmitter accuracy:
Sensor accuracy:
Total unit accuracy:
+/-0.15%
+/-1%
+/-1.15%
Transmitter unit measurement range:
+/-62Pa
Sensor drift:
<0.25% full scale/year
Radio transmitter functions:
Radio frequency options:
900 MHz Band USA and Canada
433-434MHz Band Rest of the World
Radio power:
Max 10mW, duty cycle <0.1%
Radio range:
300m over open ground
Battery Life:
(Transmitter only) Up to 1.5 years (dependent on use)
Software required:
W900 – Standard EMS Software Package
W906 – Validated EMS Software Package
*See EMS datasheet for further options
Hardware required:
N-NR001F1 – Hanwell IceSpy Network Receiver (Rest of the World)
IN-NR001F2- Hanwell IceSpy Network Receiver (USA and Canada)
IN-RP002F1 – Hanwell IceSpy Repeater (Rest of the World)
IN-RP002F2 – Hanwell IceSpy Repeater (USA and Canada)
(Repeater only required in special circumstances which would be discussed with you following your Signal Strength test.)
Probe supplied:
J266 – Prosense 4-20mA Differential Pressure Transmitter
The Hanwell EMS Software
The Hanwell environmental monitoring software (EMS) can be used with Hanwell Pro and Hanwell IceSpy hardware to provide a secure and flexible tool for comprehensive data collection, analysis and immediate alarm notification.
Validatable software
Specifically for the pharmaceutical sectors, we provide comprehensive IQ and OQ protocols and custom protocols, where required. Trained engineers and distributors are available for smooth validation protocol execution.
EMS assists with national regulatory compliance requirements
GxP Guidelines, HACCP, FDA 21 CFR Part 11, EU GMP Annex 11, and ISO 22000 food standards
Highlighted Features
Suitable Applications
Ellab Applications
Ellab provides a wide range of diverse products and solutions for most applications, all around the world.
See all applicationsSuitable Accessories
Ellab offers a wide range of customizable fittings and essential accessories for our products. The items below are for use with this product:
Ellab Accessories
Ellab offers a wide range of customizable fittings and essential accessories for processes and studies within different industries.
See all Accessories